AK—10人工肾流动泵和负压泵的故障排除

来源 :医疗卫生装备 | 被引量 : 0次 | 上传用户:lhcllk
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
我院的两台瑞典 gambro Ak-10人工肾,投入使用三年后逐渐发生故障,并多发生在主机的流动泵和负压泵。1 故障现象接电开机后,透析压(脱水压)红色指示灯亮,报警器发出警报。2 故障查找关闭电源,拆下 AK—10主机上盖,取出前面板及负压表、温度表、电导率表头。开启电源分别发现流动泵(与主机面板垂直安装的)和负压泵(与主机面板平行安装 Two of our hospital gambro Ak-10 artificial kidneys were put into operation gradually after three years of failure, and occurred in the host of the flow pump and negative pressure pump. 1 fault phenomenon After power on, dialysis pressure (dehydration pressure) red indicator light, the alarm alarm. 2 Fault Finding Turn off the power, remove the upper cover of the AK-10 and remove the front panel and negative pressure gauge, thermometer and conductivity meter. Turn on the power to find the flow pump (mounted vertically with the host panel) and negative pressure pump (parallel to the host panel installation
其他文献
故障现象:在DIRECT和STEP方式下CRT二条波形显示正常,在STORE和AVERAGE方式下CRT第一条波形显示正常,第二条波形显示严重失真幅度过大。分析与检修:1500型肌电图机在不同显
课堂教学的质量与效率直接影响着培养目标的素质与能力。而改革课堂教学模式关键要让学生在教学过程中转变以往的学习方式,自主地发现问题、提出问题并解决问题。我在教学过
擦澡是卧床病人生活护理中不可缺少的一项重要工作。但擦澡的常规操作较麻烦,且病人总是得不到彻底的擦洗,缺乏盆浴的舒适感。为了解决卧床病人的盆浴问题,我院自1989年以来
中山市商业文化博物馆(香山文化博物馆)位于中山市孙文西路商业步行街,总用地面积为1050m2,建筑占地面积为798m2,总建筑面积为2124m2,为中国首家商业文化专题博物馆,共分三层
曹思源,北京学者,涉猎政治、经济、法律等领域,有“曹破产”、“曹私有”、“曹宪政”之称。1980年首先倡导破产淘汰理论,1985年出任国务院破产法起草工作小组组长,主持起草
据相关史料研究,唐宋时期的“茶药”就是茶,其形制除团饼茶和散叶茶外还有丹丸形的,如当时径山茶就应该是这种丹丸形的茶药,并用这种茶举行径山茶宴。宋代的吃茶方法也是丰富
上个月我去西安出差,闲暇之余,我到西安的小吃街品尝了几回羊肉泡馍,那种香而不腻的感觉给我留下了美好的印象。于是我想,这羊肉泡馍肯定有点来历。我怀着好奇心,找到过去曾
第1,2,3,10题为填充题,只要求写结果,不要求写解题过程.其他各题要求解题过程.1.图1.1所示平衡系统中,物体Ⅰ,Ⅱ,Ⅲ,Ⅳ之间分别通过光滑铰链 A,B,C 连接.O,E 为固定支座,D,F
近日,中国国家标准化管理委员会主任李忠海指出,要把入世对中国企业和国家的不利影响降低到最低限度,当务之急是必须充分利用5年的缓冲期,把企业主导产品按照国际先进标准组织生产
我院一台上海医用分析仪器厂生产的GD811型连续式比色计,由于具有连续吸液、排液、连续进样等优点而受检验人员的欢迎。但使用三年多后,发生一例故障:打开电源开关,仪器正常