SJN-2200型B超维修三例

来源 :医疗装备 | 被引量 : 0次 | 上传用户:kunming
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
SJN—2200型B超是无锡海鹰集团的产品,可配用35MHz,5MHz两种线阵探头,可以获得较好的诊断图象,价格低,在基层医院有较高的普及率,现把该机的三例维修经验介绍给广大同行供参考。故障现象一:开机后,电源指示灯亮,监视器无光栅,但开、关瞬间,监视器能闪开。故障分析与 SJN-2200 type B ultrasonic is the product of Wuxi Haiying Group. It can be equipped with 35MHz, 5MHz two linear array probes, which can obtain better diagnostic images. The price is low, and it has a higher penetration rate in grassroots hospitals. The machine’s three cases of maintenance experience were introduced to the majority of their counterparts for reference. Symptom 1: After powering on, the power indicator light is on and the monitor has no light barrier. However, the monitor can be turned on immediately after turning on and off. Failure analysis and
其他文献
故障现象一:该机只能做头部和腰椎扫描,而体部病人不能扫描,显示屏提示:1,Mcuerror#2002Scannernotincworccwposition2,Mcuerror#4769MetatimeoutonendofBeammotion故障分析及排除:从以上故障现象和错 Symptom 1: The machine can only do head and l
故障现象:透视正常,点片时ready灯亮,且传片曝光程序动作正常,所点片子是白板,只在右下角病人卡处有黑化度记录。故障分析与检修:1排除洗片机故障在其它机器上所点片子冲洗后正常,
F - 8 2 0血球计数仪常规故障为小孔堵塞 ,及时快速地处理好此类故障不仅节约了时间 ,节约了资金 ,而且保障了仪器的正常使用并提高了测量的准确度。本文详尽地介绍了小孔堵
F108—V型遥控透视X线机是由北京X线机厂生产,为组合式全波整流,全封闭式X射线管头,配有日本东芝高分辨影像增强器和航天工业部研制的电视系统。现将在使用过程中出现的典型故障介
BJ—6B直线加速器具有自动楔形系统,它的60度楔形板在所给信号的控制下,自动进入或退出照射野。根据WEDGE野和OPEN野照射剂量的不同比例分配,可以合成60—0度之间的任何楔形板。OPEN野时60度
例一:用户反应一台开放式新生儿抢救台超温报警。分析检修:(1)该机工作温度设定在26℃,而显示屏上为283℃。此机在偏离设定温度2℃就会自动报警,根据现象可判断报警部分无故
例一:阿洛卡256B超机显示器故障故障现象:显示器高亮度、亮度失控、满屏出现回扫线。故障检修:根据故障现象,重点检查CRT的阴极电压和栅极电压,测阴极直流电压为+80V,属正常;检查
故障现象:机器开机后自检失败,出现“F16Pos pressure”报警.rn