溶出仪自动取样系统的校准

来源 :中国现代应用药学 | 被引量 : 0次 | 上传用户:TomasZhang_888
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
目的对溶出仪自动取样系统进行校准。方法采用经机械校准合格的溶出仪,对D、K 2种自动取样系统的取样体积、取样时间进行校准,对溶出仪采用自动与手动取样时的整机性能进行校准,并比对连续多个取样时间点自动与同时手动取样的水杨酸片的溶出量以计算偏差。结果溶出仪机械性能校准合格,自动与手动取样时溶出仪的整机性能校准合格,D、K 2种自动取样系统体积误差SS≤±2%,精密度SR≤3%,取样值滤过时间均<30 s,表明取样快速,取样体积准确、精密;D自动取样系统与手动取样溶出的比值R的均值在篮法和桨法下均<100%,且有部分点<95%,而K自动取样系统的R均值约为98%~102%,表明D系统存在一定偏差,而K系统无明显偏差。结论应对自动取样系统进行校准,对其偏差进行评估,制定可接受标准,建议体积误差SS≤±5%,精密度SR≤5%,取样值滤过时间均<30 s;溶出量≤50%的点R均值为95%~105%,R的相关标准偏差≤5%;溶出量>50%的点R均值为97%~103%,R的相对标准偏差≤5%。 Objective To calibrate the dissolution instrument automatic sampling system. Methods The sample volume and sampling time of D and K automatic sampling systems were calibrated by using a qualified dissolution apparatus. The performance of the dissolution apparatus was calibrated automatically and manually, Sampling time points automatically and simultaneously manually sampled salicylic acid tablets to calculate the amount of dissolution. Results The mechanical properties of the dissolution apparatus passed the calibration. The performance of the dissolution apparatus was qualified both automatically and manually. The volume errors of D and K automatic sampling systems were SS≤ ± 2% and the precision was SR≤3%. The filtration time All <30 s, indicating that the sampling was rapid and the sampling volume was accurate and accurate. The average value of the ratio R between the D automatic sampling system and the manual sample dissolution was <100% under the basket method and the paddle method, with some points <95% The average value of R in the automatic sampling system is about 98% -102%, indicating that there is a certain deviation in the D system and no obvious deviation in the K system. Conclusion The automatic sampling system should be calibrated, the deviation should be evaluated and the acceptable standard should be established. The recommended volume error is SS≤ ± 5%, precision SR≤5%, sample filtration time <30 s, dissolution rate≤50% The mean R of the spots was 95% -105%, the relative standard deviation of R was ≤5%, the average of the spot R of dissolution> 50% was 97% -103%, and the relative standard deviation of R was ≤5%.
其他文献
针对红外焦平面器件像素间距不可能太小,从而导致热成像系统的空间分辨率不够高这一缺陷,提出一种能实现焦平面像素倍增的双向邻域外推虚拟电子微扫技术,该方法不但解决了因
会议
目的建立梯度洗脱高效液相色谱法测定利格列汀原料杂质的方法。方法色谱柱为Agilent C18柱,流动相A为磷酸盐缓冲液(取磷酸二氢钠2.0 g,加水1 000 m L溶解,并用磷酸调节p H值
本文首先介绍了国内外光纤环圈绕制技术的发展状况,从中能比较直观地看出国内与国外先进国家在光纤环圈绕制技术上的差异.同时,概述了四极对称缠绕法的工艺过程以及进行环圈
本文就贵金属纳米粒子的光学性质,主要是消光光谱特性的相关理论与方法进行了较系统的研究.理论上从最近才发展出来的数值解的方法即离散偶极近似进行研究.实验上分析了金、
失效分析应从"设计是主导:材料是基础:工艺是保证:使用是监护"等几个方面来进行综合分析.本文以钢板弹簧在产品和工程上的应用为例,阐述了上述的观点.
本文在概括碳钢弹簧钢丝的主要标准及重要术语的基础上,比较了国内外的7个弹簧钢丝标准,提出现行国标存在不足,推荐采用ISO和JIS等标准.选择弹簧钢丝时应重视区别动载弹簧和
本文通过对摩托车减震弹簧φ6.0×230断裂件的疲劳强度分析,断裂口的宏观、微观分析和测定,得出该减震弹簧早期疲劳断裂的原因是钢丝表面存在质量缺陷,并在工艺上提出了相应
本文对恶性胸膜间皮瘤的影像学诊断进行了研究。文章围绕恶性胸膜间皮瘤的发病状况、国际恶性胸膜间皮瘤(IMIG)TNM分期系统和恶性胸膜间皮瘤的影像学诊断等进行了论述。
大多数设计人员在设计组合碟形弹簧时,往往会忽视摩擦力对组合碟形弹簧特性的影响。他们按理论设计碟形弹簧的各项参数,得出理论的特性曲线。然而在最终试验时,得出的特性曲线与
本文介绍了汽车悬架簧和气门簧用线材的质量要求和国内外先进生产技术,产品实物质量和发展方向,介绍了鞍钢弹簧钢线材开发的现状和前景.