应用自制兔抗NT血清研制成NT-RIA试剂盒。抗血清特异性强,与五种多肽无交叉反应。本盒最小检出值为10.48pg/ml,标准曲线范围20~1620pg/ml。批内和批间变异系数平均为10.25%和15.30%。临床初步应用结果显示对脑梗塞和阡癌的诊断 一定意义。
NT-RIA kit was developed using homemade rabbit anti-NT serum. Antiserum is specific and has no cross-reaction with five peptides. The minimum detectable value of this kit is 10.48pg / ml, the standard curve range is 20 ~ 1620pg / ml. The average coefficient of variation within and among batches was 10.25% and 15.30%. Preliminary clinical application results show that the diagnosis of cerebral infarction and cancer is of some significance.