洪涝灾害引发医疗仪器设备故障检修实例

来源 :医疗装备 | 被引量 : 0次 | 上传用户:aolongjiutian
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
98年 6月闽江流域发生了一场百年不遇的特大洪水 ,给国家和人民的生命财产造成很大的损失。我院有部分医疗设备被淹 ,从而引发一些仪器设备发生故障 ,现将故障的排除方法介绍如下 ,供同行参考。1 自动恒温细菌培养箱出现的故障两台自动恒温细菌培养箱被水淹过 ,经 1周时间自然 In June 1998, a catastrophic flood occurred in the Minjiang River Basin for a hundred years, causing great losses to the lives and property of the country and its people. Some medical equipment in our hospital was flooded, causing some equipment and equipment to malfunction. Now we will introduce the troubleshooting methods as follows for reference by peers. 1 Failure of Automatic Thermostatic Bacteria incubator Two automatic thermostatic bacterial incubators flooded through the water for 1 week