我厂有块XCT111动圈温度指示调节仪存放一段时间后出现了故障,其现象是,仪表通电2分钟后,表内潮气遮住观察刻度的玻璃,拆开表壳,发现变压器发热,检查热电偶无问题。连续通电观察并测量变压器,结果温升达71℃,测量变压器33V端仅有几伏电压,显然是因短路造成变压器温度升高。查控温线路发现断偶线路中2CZB2二极管正反向均通,为此更换该二极管,故障被排除。此表故障是因仪表存放在电镀车间的恶劣
I plant a block XCT111 moving coil temperature indicator regulator stored for a period of time after the failure, the phenomenon is that 2 minutes after the instrument power, the table moisture cover the glass to observe the scale, open the case and found the transformer fever, check the thermoelectric Even no problem. Continuously energized to observe and measure the transformer, the resulting temperature rise of 71 ℃, measuring the voltage transformer 33V only a few volts, apparently due to short circuit caused by the transformer temperature. Check the thermostat found broken line even 2CZB2 diode positive and negative are connected, for the replacement of the diode, the fault is ruled out. This watch failure is due to the instrument stored in the electroplating workshop of the harsh