论文部分内容阅读
引言美国COBE血透机由高频电疗机干扰引起的误报警故障是罕见的,它是我院器械科建立以来,也是我从事修理工作20多年来首次所见。以前在报刊、杂志上从未见到过,安装机器时也未见文件中提到这方面有什么技术要求,而故障偏偏又在同一台机器上间断地出现。因此,当时既觉得十分意外,又感到事出有因。保修期间, COBE公司的维修中心曾查
Introduction COBE hemodialysis machine by the high frequency electrotherapy machine interference caused by false alarm failure is rare, it is my hospital machinery department since its establishment, but also my first time in more than 20 years to see the repair work. Previously in the newspapers, magazines have never seen before, when installing the machine also did not see any technical requirements mentioned in the document, but the fault happened again in the same machine intermittently appear. Therefore, I felt both very unexpected and felt a cause for trouble at that time. Warranty period, COBE’s service center has checked