本文报导了陕西省6种类型医院,9种类型检查和5种容量x线机的7095受检人次的皮肤吸收剂量,监测结果:胸透检查为10.1mGy,群体检查为7.0mGy,胆囊造影为34.5mGy,胸片为1.5mGy,四肢片为2.3mGy,头卢片为13.5mGy,腰椎片为15.0mGy,消化道检查为40.6mGy,妇女节育环检查为10.0mGy。
This article reports the skin absorption dose of 7095 subjects in 6 types of hospitals, 9 types of examinations and 5 types of x-ray machines in Shaanxi Province. The monitoring results were as follows: chest X-ray examination was 10.1 mGy, group examination was 7.0 mGy, 34.5 mGy, 1.5 mGy for the chest radiograph, 2.3 mGy for the extremities, 13.5 mGy for the first lumen, 15.0 mGy for the lumbar spine, 40.6 mGy for the digestive tract, and 10.0 mGy for the women’s IOP.