Objective:To assess the influence of mimic cardiac rate on hydrodynamics of the different mechanical prosthetic cardiac valves. Methods: US-made CarboMedics bileaflet valve and China-made Jiuling bileaflet valve and C-L tilting disc valve have been tested in a pulsatile flow simulator in the aortic position. The testing condition was set at the mimic cardiac rate of 55 beats/min,75 beats/min,100beats/min and a constant mimic cardiac output of 4L/min. The mean pressure differences (△P) ,leakage volumes (LEV) and closing volumes (CLV) across each valve,and the effective orifice areas(EOA) have been analyzed. Results:Within the range of physiology,the △P,LEV and CLV were falling as the increasing of mimic cardiac rate,and the extent of variance was larger. The EOA was increasing with the increase of the mimic cardiac rate. It is a different response as the altering of the cardiac rate for the different type of the mechanical prosthetic cardiac valves.Conclusions:The change of the mimic cardiac rate can affect the hydrodynamics of the mechanical prosthetic cardiac valves. The hydrodynamics of the bileaflet valve prosthesis is better than the tilting disc valve.