论文部分内容阅读
270°偏转磁铁系统是 14 MeV医用电子驻波直线加速器上的关键部件。通过用 TRANSPORT—EM/PC程序和POISSON程序计算,完成了偏转磁铁样机设计方案。270°偏转 磁铁样机与加速管联合进行了出束试验,测试结果表明,具有 6~14 MeV能量范围、不同的初 始发射度的电子束流,通过270°偏转磁铁系统受到消色散和聚焦作用,传输到靶点在能量15.1 MeV时,束流强度为3.5mA,束流斑点直径约为2mm。上述结果证明270°偏转磁铁系统性能完 全满足 14 MeV医用直线加速器放射治疗条件要求。
The 270 ° deflection magnet system is a key component of the 14 MeV medical electron standing wave linear accelerator. By using the TRANSPORT-EM / PC program and the POISSON program, the design of the deflection magnet prototype was completed. The beam deflection test was carried out with a 270 ° deflection magnet prototype and an accelerating tube. The test results show that electron beams with different initial emittance in the energy range of 6-14 MeV are subjected to the decoloring and focusing effect by the 270 ° deflection magnet system, The beam delivered to the target at a energy of 15.1 MeV had a beam intensity of 3.5 mA and a beam spot diameter of about 2 mm. The above results demonstrate that the performance of the 270 ° deflection magnet system fully meets the radiotherapy conditions for 14 MeV medical linear accelerators.