BJ—4直线加速器特殊故障一例分析检修

来源 :医疗装备 | 被引量 : 0次 | 上传用户:chendegeng1234
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
故障现象:开机,机器CGY/MIN数码表不时出现-400V并报警,无法正常工作,按复位键无效。分析检修:BJ—4剂量显示过程如下,当X线通过加速器平板电离室时就要产生电离电流。该弱电流(约10-9A~10-7A)经过K放大器积分转换成50mV-5V的电压 Symptoms: After booting, the machine CGY/MIN digital watch appears -400V from time to time and alarm, can not work, press the reset button is invalid. Analytical Maintenance: The BJ-4 dose display process is as follows. When the X-ray passes through the accelerator plate ionization chamber, ionization current is generated. The weak current (about 10-9A ~ 10-7A) is converted into a voltage of 50mV-5V by K amplifier integration
其他文献
故障现象监护仪无创血压(NIBP)测量充气结束时,压力值显示较正常值低约10kPa,并出现E17故障报警。在按SILENCE/RESET键时使其复位,再次测量血压,故障现象同上。故障原因分析E17故障是指袖带充气时间超过60秒未达所
Some people believe that humans could live on the moon in the future.That is because our planet,the earth,is too crowded and polluted.The moon is another ideal
故障1 开机面板显示正常,预热2~3分钟后点灯,氪灯不能点燃,电流表电压表读数无变化,显示板显示错误信息,报警。重复点灯多次故障依旧。故障分析与检修 按常规经验,点灯时,未
 
所谓活动营销,就是在整个营销过程中,选择一个或是几个关键性的节点,安装“发动机”、“助推器”,以便给一个特定的产品做品牌建设或销售工作全力加速。 The so-called acti
医院改革只有靠挖掘医院内部潜力,强化医院科学化、规范化管理,提高医疗质量,改善服务态度,拓展医疗服务范围,增加便民措施和提高质量效益上下功夫,炼内功,从而以一流的质量,优质的服
文题阅读下面的文字,写一篇不少于800字的文章。据英国《每日邮报》报道,近日,西班牙一场自行车赛,一位名叫艾斯特万的选手距终点300米时遭遇爆胎。然而这是关系到铜牌之争的
数学对学生而言是一门基础性的学科,在学习的过程中将会涉及到很多关于符号、概念的知识点。但是学生在学习的过程中往往以形象思维为主,所以对学生而言,学习数学是有一定的
近年来由于经济发展势头强劲,贸易持续顺差,中国因此成为众多西方国家反倾销调查的对象,并通过对我国“非市场经济国家”的认定,确定倾销成立,征收高额反倾销税。因此认清“
故障现象:仪器通电开机后,由去颤器机壳内部传出一种异常的噪音及飘出一股焦糊气味。检修步骤:打开去颤器外壳,发现噪音来自JK4高压继电器内部且金属外壳亦发热烫手,高压继电器JK3线