故障现象按ON/OFFCHARGE键 ,显示F-38。故障检测根据故障代码表提示 ,F -38故障出于输液管道检测电路。应检查CN803的连接情况 ,必要时更换FSR。Baxter输液泵为方便维修 ,设置了自动测试模式 ,在自动测试模式下 ,可做输液泵校正、LCD和LED显示检查、蠕动泵和阻塞传感器检查、气
Symptom Press ON/OFF CHARGE key to display F-38. Fault detection According to the fault code table, the F-38 fault is detected by the infusion line detection circuit. Check the connection of the CN803 and replace the FSR if necessary. The Baxter infusion pump is equipped with an automatic test mode for easy maintenance. In the automatic test mode, the infusion pump calibration, LCD and LED display check, peristaltic pump and obstruction sensor check, and gas can be performed.