现将深圳“康立AFT—400”电解质分析仪几年来发现的使用说明书以外的常见故障的排除方法报告如下,供同行借鉴。1电极方面1.1定标不易通过或测出的某离子结果前带(。)。原因:电极漂移。可自动或手工活化电极(根据电极漂移情况选用清洗液或新鲜血清作活化剂),反复冲洗。
Now the Shenzhen “Kangli AFT-400” electrolyte analyzer found in the use of manual failures in addition to common troubleshooting methods reported below, for peer reference. 1 Electrode aspect 1.1 Calibration is not easy to pass or measure the results of a certain ion before (.). Cause: The electrode drifts. The electrode can be activated automatically or manually (using a cleaning fluid or fresh serum as the activator depending on the electrode drift) and rinsed repeatedly.