目的探讨胸部CT扫描参数,控制X射线辐射剂量。方法随机选取来我科做胸部CT复查的30个病例,体重指数在中等以下(男性平均23.09,女性平均19.86),使用低剂量(20 mAs)扫描,扫描图像由两位有经验的主治医师盲氏阅片,评判低剂量和自身常规剂量(100 mAs)扫描图像中左上肺血管分辨率,将图像分成“好”、“一般”、“差”三级。结果①低剂量扫描的图像“好”59份,“一般”1份,无“差”级图像,图像质量全部符合诊断要求;②胸部常规剂量扫描容积CT剂量指数(CTDIvol)平均5.89 mGy,低剂量扫描CTDIvol平均1.22 mGy,低剂量扫描的辐射剂量只有常规剂量扫描的大约1/5。结论在体重指数为中等以下时,胸部CT复查低剂量扫描的图像可以满足诊断要求;低剂量扫描的辐射剂量只有常规剂量扫描的1/5。
Objective To investigate the chest CT scan parameters and control the dose of X-ray radiation. Methods Thirty cases were randomly selected from our department for chest CT. The body mass index was below average (23.09 for males and 19.86 for females), and low-dose (20 mAs) scans were performed. The images were blindly reviewed by two experienced physicians ’S interpretation of low-dose and its own conventional dose (100 mAs) scan the image of the left upper pulmonary vascular resolution, the image is divided into “good ”, “general ”, “poor ” three. Results ① The images of low-dose scan were 59 in “good”, 1 in “normal”, and no in “poor” in images, all of them were in accordance with the diagnostic requirements; ② CT dose index ) Averaged 5.89 mGy, with an average of 1.22 mGy for low-dose CTDIvol and a low-dose scan of about 1/5 of the normal dose. Conclusions When the body mass index is below the middle level, the chest CT scan can meet the diagnostic requirements for low-dose scanning images. The dose of low-dose scanning is only 1/5 of the conventional dose scanning.