AD-260稀释器吸量减少故障排除一例

来源 :医疗装备 | 被引量 : 0次 | 上传用户:xiaoyuerhaha
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
故障现象:出液量减少。分析检查:查外围管道无阻寒或漏液现象。打开机壳,开机观察吸放液凸轮运动,吸放过程购能到位,再行吸液后,拔除连接机体的进液管道,掀吸放液按键放液,见液体倒流,确定为进液口活塞故障。拆开活塞仔细观察,发现一微小结晶体,经有蒸馏水洗清除后 Symptoms: Reduced fluid output. Analysis and inspection: Check the surrounding pipeline without any cold or leakage. Open the case, start to observe the suction and discharge cam movement, the suction and discharge process is in place, and after the liquid is sucked again, remove the liquid inlet pipe connecting the body and lift the liquid and press the liquid to release the liquid. See the liquid backflow and determine it as the liquid inlet Piston failure. After careful observation of the disassembled piston, a tiny crystal body was found and cleaned with distilled water.
其他文献
贝克曼SYNCHRONEL-ISE电解质系统是美国Beckman公司最新一代电解质测定系统,首次采用间接离子选择电极法,使电极寿命大大延长;工作效率高,每小时约可测定100T样本;该电解质系统是卫
故障一:相线与零线短路,造成电压突然升高,烧毁仪器。我院的供电是采用三相四线制的。其中墙内的动力线是由一根相线与一根零线引出的220V的电源。由于医院电源进线380V的相线与
美国ASCA全自动生化分析仪是美国MI-CROGENICS仪器公司生产的一种离心式全自动生化分析仪,该机首先在中国大陆使用全汉化的菜单操作,非常适合于中国国情。该机全套包括装载仪、
故障现象:开机,描笔单偏,调移位电位器,描笔能上下大幅度移动,但很难把描笔调到中间位置。维修过程:引起描笔单偏的原因有:(1)电源(2)放大系统(3)位置反馈系统。①首先测电源电压正常,调W4
卫生部兰州生物制品研究所利用世行贷款进行疫苗生产技术改造,改、扩建冻干麻疹活疫苗、百日咳菌苗、白喉、破伤风类毒素三联制剂等三条疫苗生产线。该项目总投资近4.6亿人民
卫生部陈敏章部长于96年12月19日下午在全国卫生工作会议上所做的《深化改革,加快开发,开创卫生工作新局面》的报告中指出今后要努力抓好七个重点领域的工作。各级党委、政
故障现象:开机后,按“COUNT”键,负压系统工作正常,水银能正常复位,计数器和计时器均能正常工作,但到检测时间(约13秒)时显示屏仍然继续计数、计时不能停止,无法得出测量结果。原理分
高胆汁酸血症出现于多种肝脏疾病。几乎在所有肝胆疾病中,急性病变的胆汁酸异常率高达100%,慢性病变的肝胆疾病血清胆汁酸异常达92%。当胆酸迅速上升超过肾脏的排泄功能,病人
基层医院地处农村,收治的住院病人大多是农民,病房内陪护多,病人卫生习惯不好,文化层次低。再加上医院基本设施不全,床单元配备不齐,护士对病房管理重视不够,因此基层医院的
现在医学领域中所用的X线摄影唁金及胶片的尺寸有5×7、8×10、10×12×11×14、12×15、14×14、14×17英寸共七种规格,为了使摄影唁金及胶片的尺寸与差异不等的体位设计合理