目的 对自行研制的立位耐力监测系统的稳定性和可行性进行评价。 方法 试验对象为 6 1名健康飞行人员。以常规主动立位耐力试验为标准方法 ,应用立位耐力监测系统与常规主动立位耐力试验进行心率、收缩压、舒张压和平均动脉压监测 ,对各生理参数进行对比分析。立位耐力监测系统是由计算机控制监测台和自动记录包括有 10余项生理参数和图形的立位耐力评价系统。 结果 经校正后 ,两种方法在立位即刻心率分别为 ,立位耐力监测系统 76 .6 9± 13.19,常规主动立位耐力试验 88.2 0± 11.6 3,两者有非常显著性差异 (P<0 .0 1)。在第 2 0 m in舒张压、平均动脉压分别为 :立位耐力监测系统 (77.78± 10 .6 5 ) mm Hg、(89.78± 11.5 5 ) mm Hg,常规主动立位耐力试验 73.80±11.10 ,85 .35± 8.93,两者有显著性差异 (P<0 .0 5 ) ,其余两种方法各时间段的心率、收缩压、舒张压和平均动脉压相比 ,均无显著性差异 (P>0 .0 5 ) ,而且 ,两种方法的线型趋势图几近重合。 结论 立位耐力监测系统稳定、可行 ,可以替代常规主动立位耐力试验。
Objective To evaluate the stability and feasibility of self-developed stand-endurance monitoring system. Methods The subjects were 61 healthy flight personnel. Taking the routine active standing endurance test as the standard method, the monitoring of heart rate, systolic pressure, diastolic pressure and mean arterial pressure were performed by using the vertical endurance monitoring system and the conventional active standing endurance test, and the physiological parameters were compared. Stand endurance monitoring system is a computer-controlled monitoring station and automatic recording includes more than 10 physiological parameters and graphics stand endurance evaluation system. Results After adjustment, the instantaneous heart rate of the two methods were 76.69 ± 13.19 in standing position endurance test system and 88.2 ± 11.63 in active stand-up position test, respectively. There was a significant difference between the two methods (P < 0 .0 1). The diastolic blood pressure and mean arterial pressure at 2 0 m in were respectively: stand-up endurance monitoring system (77.78 ± 10.65) mm Hg, (89.78 ± 11.5 5) mm Hg, conventional stand-alone endurance test 73.80 ± 11.10, 85.35 ± 8.93, respectively (P <0.05). There was no significant difference in heart rate, systolic pressure, diastolic pressure and mean arterial pressure between the other two methods (P > 0 .0 5), Moreover, the linear trend of the two methods almost coincide. Conclusion The stand endurance monitoring system is stable and feasible, which can replace the routine active stand endurance test.