在色谱操作过程中,鉴定器有时受固定相流失及样品中的高沸点成分、易分解及腐蚀性物质的作用而被沾污,以至不能正常进行工作,因而提出了如何清洗鉴定器的问题。若沾污的物质仅限于高沸点成分,通常可将鉴定器加热至最高使用温度后,再通入载气,就可清除。使用有放射源的鉴定器时加热要多加小心,例如通常以氚源作成的电子捕获鉴定器一般都不能超过200度,此外还应注意加热的温度不能损坏鉴定器的绝缘材料。如用加热法不适宜,也可以用纯的丙酮等溶液从进样口注入(每次可注入几十微升)进行清洗,这在沾污程度较轻时是有效的。
During the chromatographic operation, the identifier is sometimes contaminated by the loss of the stationary phase and the high-boiling components in the sample, which are easily decomposed and corrosive substances, so that the normal operation can not be performed. Therefore, the problem of how to clean the identifier is proposed. If the contaminated substance is limited to high-boiling components, it is usually removed by heating the detector to the highest temperature and then feeding the carrier gas. Care should be taken when using detectors with radioactive sources. For example, electron capture detectors, usually made from tritium sources, should generally not exceed 200 degrees. In addition, care should be taken that the heating temperature does not damage the insulation of the detector. If the heating method is not suitable, you can also use pure acetone and other solutions injected from the inlet (each injection can dozens of microliters) for cleaning, which is less contaminated when the effective.