百特550肾机低电导报警紧急处理一例

来源 :医疗装备 | 被引量 : 0次 | 上传用户:D243435dsfsdfqwe
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
故障现象:肾机开机后一直低电导报警,有时间HiCond键能消除低电导报警,但工作一段时间又出现低电导报警。分析与检修过程:引起低电导报警的原因有很多:如配制的浓缩液有问题;比例泵故障;电导监测电路有故障及电导电池损坏等。由于我院有三台百特550肾机,而其 Symptom: After the kidney machine is turned on, it has been low-conductivity alarm. There is a time when the HiCond key can eliminate the low-conductivity alarm, but a low-conductivity alarm occurs after working for a period of time. Analysis and Maintenance Processes: There are many causes of low conductivity alarms, such as problems with formulated concentrates, failure of proportional pumps, faults in the conductivity monitoring circuit, and damage to the conductance battery. Since there are three Baxter 550 kidney machines in our hospital,
其他文献
食管巨型锐利性异物临床上并非罕见病例,用惯用方法取出,既易损伤食管又可能发生危及生命的并发症,而开胸、颈侧过路取异物又绝非上策,为此我们自行设计制作了一种食管保护装置,暂
萆始载于《本经》 ,被列为中品 ,功能利湿浊 ,祛风湿 ,主治膏淋、白浊、带下疮疡、风湿痹痛。笔者将其用于治乳糜尿、肾炎、肝硬化等病 ,效果甚好。试举验案 3则 ,供同道参
DFT-1型肺通气量仪是一种电子自动化程度较高、接收灵敏度较强的自动化检测仪器。适用于体检及肺功能的检测。故障现象肺通气仪探头失灵分析检修由于该探头灵敏度较高,很小的气流就
故障现象:扫描中GANRY突然停止在12:00的位置。手动控制和计算机控制均失灵。ODC台上的CRT上显示“STARTSCANSEQEBCEER-ROR”,“STRSCNCALLERROR”。故障检修与分析:在首先测量床台主电源的工作电压+24V、+5V Symptom: The GANRY suddenly st
近年来,随着科技兴农意识的增强,各地农村纷纷建立了“庄稼医院”,这对指导农民生产、夺取农业生产丰收起到了很好的作用。然而,笔者在下乡走访时了解到,有一些“庄稼医院”名不副
抽取胸腹腔的积液,是治疗造成胸腹腔积液疾病的治疗手段之一。常用的方法是采用100ml注射器多次进行抽吸,抽出的液体推入溶液盆中,带有结核杆菌或肝炎病毒的胸腹腔积液经常会溅
由北京大学图书馆和北京高校图书馆期刊工作研究会共同组织评选的第四届“新闻核心期刊”,近日通过评审鉴定,《现代传播》再次当选。
香港WEX公司生产的HD-3L型心电多功能自动诊断仪是一台集计算机系统、人体生理信号采集处理系统、数据处理分析系统于一体的智能化医疗设备。当今智能化医疗设备已是医疗设备领域的主
现代科学技术,促进了医疗设备的高速发展。近几年来,大量高精尖设备进入医院,创造了极大的社会效益和经济效益,对保障人民健康和加快医疗卫生事业的发展起了重要作用。如何保证这
自动洗片机工作效率高,质量稳定,清晰度好等优点被各大中医院普遍使用,暗室工作逐渐由手工操作变为机械操作。我科先后引进国产、进口洗片机数台,大大减轻暗室劳动强度,提高了工作