目的了解浙江省消毒检测实验室质量控制水平,提高全省疾控系统和医疗机构对消毒剂性能检测能力。方法采用理化分析方法,选部分疾病预防控制中心和医疗机构对戊二醛消毒液有效含量检测能力进行调查。结果15所疾病预防控制中心实验室对两种浓度(含量18 g/L与16 g/L)的戊二醛消毒液检测报告的结果符合率为86.7%。26家医疗机构对含量18 g/L戊二醛消毒液检测结果符合率为69.2%,对含量16 g/L戊二醛消毒液检测结果符合率为26.9%。结论浙江省医疗机构应进一步加强戊二醛消毒剂的正确使用;疾病预防控制中心中尚未开展戊二醛有效含量测试的单位,应尽早建立相应的测试方法,加强对医院使用中消毒剂的质量检测。
Objective To understand the quality control of disinfection testing laboratory in Zhejiang Province and improve the capability of testing the performance of disinfectants in disease control systems and medical institutions across the province. Methods Physical and chemical analysis methods were used to select part of the disease prevention and control centers and medical institutions to investigate the detection of the effective content of glutaraldehyde disinfectant. Results The results of 15 reports of glutaraldehyde disinfectant in the laboratories of two Centers for Disease Control and Prevention in both concentrations (18 g / L and 16 g / L) were 86.7%. In 26 medical institutions, the coincidence rate of glutaraldehyde disinfectant with 18 g / L disinfectant was 69.2%, and the coincidence rate with glutathione disinfectant with 16 g / L disinfectant was 26.9%. Conclusion Zhejiang medical institutions should further strengthen the proper use of glutaraldehyde disinfectant; CDC units that have not yet carried out the glutaraldehyde test should establish test methods as soon as possible to enhance the quality of disinfectants used in hospitals Detection.