论文部分内容阅读
当前,国产人工肾机,有的未设置漏血监护仪,有的虽设置但性能较差,不能及时报警。为提高漏血的监视性能,使医护人员能及时采取措施,浙江医科大学附属第一医院泌尿外科透析室和浙江医科大学物理学教研室共同协作,应用红外线监测的新技术,试制成功ZY—820型红外线漏血监护仪。在完成实验室血液检测后,用于临床作血液透析监视,效果良好,性能优于可见灯光感应的漏血监视,其监视渗漏血量在超过设定万分之五时,即能在2秒钟内正确显示声、光警报信
At present, domestic artificial kidney machine, and some did not set blood leakage monitor, although some settings but the performance is poor, not timely alarm. In order to improve the monitoring performance of blood leaking so that medical staff can take timely measures, the First Affiliated Hospital of Zhejiang Medical Urology dialysis room and the Zhejiang Medical University Physics Department work together to apply infrared monitoring of new technologies, successful trial ZY-820 type Infrared blood leakage monitor. After the completion of the laboratory blood test for clinical hemodialysis monitoring, the effect is good, the performance is better than the visible light-induced leakage monitoring, monitoring the leakage of blood in more than set five ten thousandths, that is, in 2 Sound correctly within seconds, light alarm