在定量分析中,经常遇到同一试样在含量线性改变过程中,出现反线性数据现象。这纯粹是由误差造成的.在分光光度计的使用上,较大的误差的常常成组地出现,影响着数据分析.造成这种误差的原因之一,就是波长设定不准确,以致所测得吸收度 A≠A_(amx)。本文试以在所选波长附近一微小范围内,调整波长在 A_(max)所在位置,以排除误差的影响.1 仪器与试药:UV-754紫外分光光度计(上海);重蒸馏水,人民医院制剂室新制;维生素 B_1,哈尔滨制药三厂
In quantitative analysis, often encounter the same sample in the process of linear changes, the phenomenon of anti-linear data. This is purely caused by errors.Of the use of spectrophotometers, large errors often appear in groups, affecting the data analysis.One reason for this error is the wavelength setting is not accurate, so that the Measured absorbance A ≠ A_ (amx). In this paper, a small range of wavelengths near the selected wavelength, the adjustment wavelength in the location of A max, in order to exclude the impact of error.1 Instruments and reagents: UV-754 UV spectrophotometer (Shanghai); Double distilled water, the people Hospital preparation room new system; vitamin B_1, Harbin Pharmaceutical Factory