论文部分内容阅读
日立P-C——125H型X线机,1989年投入使用,近10年来没有出现大故障。今年3月,在给病人做导管介入治疗时,显示器突然无图像,变黑。经检查,系影像增强器的摄像机系统损坏。经过分析,最后判断为摄像机高压包T201短路烧坏。曾试验仿制未成,经过调查国内无此配件,和日立公
Hitachi P-C - 125H X-ray machine, put into use in 1989, no major failure in the past 10 years. In March of this year, when the patient was given a catheter intervention, the monitor suddenly had no image and turned black. Upon inspection, the camera system of the image intensifier was damaged. After analysis, the final judgment is that the camera high-voltage package T201 short-circuit burnout. Tested imitation was not completed, after investigation, there is no such accessory, and Hitachi