该机是原上海医疗仪器厂现上海医用核子仪器厂生产的一种单焦点、半波自整流组合机头式X线机,可用作一般透视、摄影。改进后的F30-Ⅱ型X线机附有床下活动滤线器,并在荧光屏上装有简单的适时摄影装置。整机结构简单,安装方便,是公社卫生院的主要设备。为便利检修起见,将本机在使用中常见故障及其维修分别介绍如下,供医务和检修人员参考。一、低压电路常见故障及检修 (1)机器不得电,电源指示灯不亮,电源电压表无指示这种故障除电源外,常见有以下两种原因:一是操纵台内电源熔断器RD烧断,检
The machine is a single-focus, half-wave self-rectifying combination head type X-ray machine produced by the former Shanghai Medical Instrument Factory, Shanghai Medical Nuclear Instrument Factory. It can be used for general perspective and photography. The improved F30-II X-ray machine is equipped with an under-bed motion filter and a simple, timely photographing device on the screen. The structure of the whole machine is simple and easy to install. It is the main equipment of the public hospitals. For the convenience of inspection, the common faults and maintenance of the unit during use are introduced as follows for reference of medical personnel and maintenance personnel. First, the low-voltage circuit common faults and overhaul (1) The machine must not be powered, the power indicator light is not lit, the power voltmeter no indication This kind of fault In addition to power, there are two common reasons: First, the control panel power fuse RD burn Break, check