在能谱测量装置中,要求光电倍增管有好的能量分辨率和好的时间稳定性。特别是测量低强度的γ光子或电子时,长时间的测量更要求光电倍增管有好的稳定性。为了建立全吸收契伦柯夫谱仪装置,我们对几种大面积光电倍增管作了调试和选择实验。实验中把3″×3″NaI(TI)晶体耦合到光阴极,用~(137)Cs源的0.661MeVγ光子照射,将光电倍增管的输出脉冲输入4096道幅度分析器,进行脉冲幅度分析。对不同类型管子的测
In spectrometry devices, photomultiplier tubes are required to have good energy resolution and good time stability. Especially for the measurement of low-intensity gamma photons or electrons, long-time measurements require a good stability of the photomultiplier tube. In order to establish a fully absorbed Cherenkov spectrometer device, we commissioned and selected several large-area photomultiplier tubes. In the experiment, the 3 “× 3” NaI (TI) crystal was coupled to the photocathode. The output of the photomultiplier tube was input into 4096 amplitude analyzer by 0.661MeVγ photon from 137Cs source, and the pulse amplitude was analyzed. Measurement of different types of pipes