采用插有起搏器脉冲分析软件的微电脑三道心电图仪,直接描记脉冲波并自动显示频率、脉宽、振幅及衰减比率等参数,用于随访埋藏起搏器病人95例,共作脉冲分析232次。脉冲分析能辅助判断电池耗竭、导线绝缘层破坏、导线断裂或与起搏器接口松脱以及起搏器线路故障,特别是脉冲波形、抑幅及衰减比率三个参数为一般随访手段所不能检测,而对及时诊断处理却有较大帮助。本方法简便快捷,有临床应用价值。
Using the microcomputer three electrocardiograph with pulse analysis software inserted in the pacemaker, the pulse wave was directly traced and the parameters such as frequency, pulse width, amplitude and attenuation ratio were automatically displayed for the follow-up of 95 patients with pacemakers and pulse analysis 232 times. Pulse analysis can help determine the battery drain, wire insulation damage, wire breakage or loosened with the pacemaker interface and pacemaker line failure, in particular the pulse waveform, amplitude and attenuation ratio of the three parameters for the general follow-up means can not detect , But on the timely diagnosis and treatment have greater help. The method is simple and quick, and has clinical value.