本文是对F78—Ⅱ型诊视床在使用过程中,由于使用者操作不谨慎,在卧床或向负角度下倾时,障碍物(如椅子等)存在,使床的侧护板等一些部件损坏;在诊视床向正90°立起时,在床身足端压有异物(如脚踏开关、木块)等,慌忙中停机不够及时,床身继续运转,结果床身足端底板损环,严重时将底座的固定螺栓拔起,床体松动造成危险,严重影响诊视床的使用。由于上述问题存在,除使用者操作不谨慎外,该诊视床保护措施不够完备也是一方面,因此,建议在该诊视床
This article is the F78-Ⅱ-type bed in the use of the bed, due to the user’s operation is not cautious, in the bed or down to the negative angle, obstacles (such as chairs, etc.) exist, the bed side shield and some other parts damage ; Bed in the bed to the positive 90 ° erected when the bed at the end of the foot pressure of foreign bodies (such as foot switch, wood), hurried down in not timely enough, the bed continues to run, the results of the foot bed end of the plate loss , Serious when the base of the fixed bolt pulled up, the bed loose caused a dangerous situation, seriously affecting the use of diagnostic bed. Due to the above problems, in addition to the user operation is not cautious, the bed protection measures are not complete, but also on the one hand, it is recommended in the bed