A method based on the combination of coprecipitation with inductively coupled plasma atomic emission spectrometry (ICP-AES) was developed for the determination of impurities in high-purity sodium tungstate. Six dements (Co, Cu, Fe, Mn, Ni, and Pb) were coprecipitated by lanthanum hydroxide so as to be concentrated and separated from the tungsten matrix. Effects of some factors on the recoveries of the analytes and on the residual amount of sodium tungstate were investigated, and the optimum conditions for the coprecipitation were proposed. Matrix-matching calibration curve method was used for the analysis. It is shown that the elements mentioned above can be quantitatively recovered. The detection limits for Co, Cu, Fe, Mn, Ni, and Pb are 0.07, 0.4, 0.2, 0.1, 0.6, and 1.3 μg.g~(-1), respectively. The recoveries vary from 92.5% to 108%, and the relative standard deviations (RSDs) are in the range of 3.1%-5.5%.
A method based on the combination of coprecipitation with inductively coupled plasma atomic emission spectrometry (ICP-AES) was developed for the determination of impurities in high-purity sodium tungstate. Six dements (Co, Cu, Fe, Mn, Ni, and Pb) were coprecipitated by lanthanum hydroxide so as to be concentrated and separated from the tungsten matrix. Effects of some factors on the recoveries of the analytes and on the residual amount of sodium tungstate were investigated, and the optimum conditions for the coprecipitation were proposed. The detection limits for Co, Cu, Fe, Mn, Ni, and Pb are 0.07, 0.4, 0.2, 0.1, 0.6, and 1.3 μg.g -1, respectively. The recoveries vary from 92.5% to 108%, and the relative standard deviations (RSDs) are in the range of 3.1% -5.5%.