日本产NiproNCU-10型血液透析机使用一段时间后易发生“流量停止异常”故障,此时按机上液晶屏提示的故障排除方法检查相关器件往往并不能解决问题。其实大多数此类故障的产生原因不是器件损坏,而是流量显示器(浮子)的玻璃外壳被铁锈脏污所引起。NCU-10型
Japan NiproNCU-10-type hemodialysis machine prone to “flow stop abnormal” fault after a period of time, then press the LCD screen prompts troubleshooting methods related devices often does not solve the problem. In fact, most of these failures are not caused by device damage, but the glass case of the flow indicator (float) is caused by the rusting of the rust. NCU-10 type