F-820血球计数仪特殊故障一例

来源 :医疗卫生装备 | 被引量 : 0次 | 上传用户:TomasZhang_888
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
根据用户主诉:送修前该机曾显示“WASTEFULL”(废液瓶满),但当时废液瓶并不满,用户认为故障报警可能是由废液瓶液面传感器固定螺母处漏气所致,所以用力紧固该螺母,因受力过大致使传感器固定螺杆(工程塑料制)断裂,造成废液液面传感器无法固定,废液瓶无负压。经向F820 According to the main complaint of the user, the machine displayed “WASTEFULL” (waste bottle full) before repair, but at that time, the waste bottle was not full. The user thinks that the failure alarm may be caused by leakage of the fixing nut at the liquid level sensor of the waste bottle. Therefore, when the nut is firmly tightened, the sensor fixing screw (engine made of plastic) is broken due to the force and the liquid level sensor of the waste liquid cannot be fixed, and the waste liquid bottle has no negative pressure. Longitudinal F820
其他文献
亚低温对缺血脑具有保护作用和缺血神经元损伤存在温度依赖性。其机制尚未完全阐明 ,涉及复杂的生物学级联反应过程 ,包括对兴奋性和抑制性递质的作用 ,对Ca2 + 超载和Glu受
卵巢恶性苗勒管混合瘤(malignant mixed mulleriantumors of the ovary,OMMMT)占卵巢恶性肿瘤的1%左右[1]。本文报道1例OMMMT合并乙型肝炎肝硬化患者术后复发,后行二次肿瘤细
ILEX电解质仪是深圳三龙公司生产的一种自动血液电解质分析仪,可直接检测血清中钾、钠、氯、离子钙,测量速度快,样品用量少,操作简便,结果准确。该仪器使用过程中出现液流过慢或堵
中华医学会的宗旨是:团结广大医学科学技术工作者和管理工作者,促进医学科学技术的繁荣和发展,促进医学科学技术的普及和推 The purpose of the Chinese Medical Associatio
第一条 为了妥善解决基本医疗保险范围以外的医疗费用问题,根据全国城镇职工医疗保险制度改革工作会议精神和《曲靖市职工医疗保险暂行办法》的有关规定,特制定本办法。第二条
BECKMANCXS全自动生化分析仪是美国贝克曼公司生产,为当今世界上最先进的生化分析仪之一,在我国使用较为广泛。我院自1996年使用后,大大提高了检测的速度和质量。对于其出现的常
医学模式转变与社会主义精神文明建设是什么关系?怎样在精神文明建设实践中促进医学模式的转变?本文试就这个问题谈点看法. What is the relationship between the transfor
故障现象:接通电源后,液晶显示状态正常,但转换导联选择开关、灵敏度选择开关、纸速选择开关等自动键盘上各项功能开关时不起作用,液晶显示情况不变化,并且没有代表选通状态的“笛
目的:探讨慢性心衰时心肌细胞肌动蛋白的变化情况。方法:采用SP免疫组化方法,对升主动脉缩窄慢性心衰大鼠心肌细胞肌动蛋白进行定性及半定量分析。结果:慢性心衰时左心室心肌细胞肌
我科于1996年购置澳大利亚TRACE公司产品CB-171半自动生化分析仪。在两年的使用过程中,发现了一些问题及处理办法。观介绍如下,以便使用本仪器的同行参考。1蒸馏水空白四光度偏