论文部分内容阅读
目的检验本实验室个人剂量监测能力。方法按照组织者比对方案要求,将多组常规使用个人剂量计交组织者,由组织者以不同能量的X、γ射线体模照射得出约定真值(参考值),由实验室测量并统计各测量值,分析测量值与参考值的相对偏差,符合组织者要求为比对合格。结果 2008~2010年连续3次比对结果合格。结论本实验室个人剂量监测质量符合国家规范相关要求。
Objective To test personal laboratory dose monitoring capabilities. Methods According to the requirements of the organizer, multiple sets of routine personal dosimeters were handed over to the organizer, and the organizer irradiated the X, γ-ray phantom with different energy to obtain the agreed true value (reference value), which was measured by the laboratory Statistics of the measured values, analysis of the relative deviation between the measured value and the reference value, in line with the requirements of the organizer for the comparison pass. Results 2008 to 2010 for three consecutive results of the comparison. Conclusion The quality of individual dose monitoring in our laboratory conforms to the relevant requirements of the national code.