过氧化氢低温等离子灭菌器使用过程中常见灭菌循环运行停止并且报警现象,这是灭菌器设计者为了避免某些可能致使灭菌失败因素存在而专门设计的报警系统,以提高灭菌的成功率。北京地坛医院自引进国外进口的STERRAD 100S型过氧化氢低温等离子灭菌器运行的5年时间内,在3 678次灭菌循环中,发生循环取消68次,发生率达到1.85%;随着使用时间延长,经过不断分析和改进操作,使得 Hydrogen peroxide low temperature plasma sterilizer common use of sterilization cycle to stop running and alarm phenomenon, which is sterilizer designers in order to avoid sterilization factors that may lead to the existence of specially designed alarm system to improve sterilization The success rate. Beijing Ditan Hospital since the introduction of imported STERRAD 100S-type hydrogen peroxide plasma sterilizer operation within 5 years, 3 678 sterilization cycles, the abolition of 68 cycles occurred, the incidence rate of 1.85%; with the use of Extended time, through continuous analysis and improvement of operations, making