我院一台上海医疗器械厂产的XG200mAX线诊断机开机后突然发生旋转阳极X线球管爆裂。现将故障分析和检修过程及今后故障的防范分述如下,希同道指正。1、故障发生过程;技照技师开机作好滤线器摄影准备,并将片盒送入片槽,这时听到旋转阳极x线球管部位内有玻璃强烈破裂声,立即
After the XG200mAX diagnostic machine manufactured by a Shanghai medical device factory in our hospital suddenly turned on, the rotating anode X-ray tube burst suddenly. The failure analysis and maintenance process and the prevention of future failures are described below as follows. 1. The process of failure occurred; technicians turned on the photographic preparation of the thread filter and sent the cassette into the slot. At this moment, there was a strong glass crack in the rotating anode x-ray tube, immediately