论文部分内容阅读
目的 :为解决 Sysm ex SF- 30 0 0血液分析仪稀释液试剂国产化 ,比较自配稀释液与进口稀释液是否有差异。方法 :通过检测进口稀释液的主要参数 ,确定自配稀释液的配方。结果 :自配和进口稀释液理化指标一致 ;它们检测 2 0份抗凝静脉血标本 ,所得七项主要指标均数相近 ,经配对资料 t检验 ,P值均大于 0 .0 5 ,相关系数均大于 0 .98;而且自配试剂重复性、稳定性良好。结论 :自配 Sysmex SF- 30 0 0血液分析仪稀释液完全可以代替进口稀释液
Purpose : To resolve the localization of Sysmex SF-300 Blood Analyzer Diluent Reagent, compare the difference between self-made diluent and import diluent. Method : Determine the formula of the self-made diluent by testing the main parameters of the imported diluent. Results: The physical and chemical indicators of the self-made and imported diluents were the same; they detected 20 samples of anticoagulated venous blood, and the seven main indicators obtained were similar to each other. After paired data t-test, the P values were all greater than 0.05, and the correlation coefficients were both It is more than 0.98, and the self-administered reagent has good repeatability and stability. Conclusion : The Sysmex SF-300 Blood Analyzer Diluent can completely replace the import diluent