论文部分内容阅读
An optical fiber bienzyme sensor based on the luminol chemiluminescent reaction was developed and demonstrated to be sensitive to glucose. Glucose oxidase(GOD) and horseradish peroxidase(HRP) were co-immobilized by microencapsulation in a sol-gel film derived from tetraethyl orthosilicate(TEOS). The calibration plots for glucose were established by the optical fiber glucose sensor fabricated by attaching the bienzyme silica gel onto the glass window of the fiber bundle. The linear range was 0 2-2 mmol/L and the detection limit was approximately 0 12 mmol/L. The relative standard deviation was 5.3% ( n =6). The proposed biosensor was applied to glucose assay in ofloxacin injection successfully.
An optical fiber biosensor based on the luminol chemiluminescent reaction was developed and demonstrated to be sensitive to glucose. Glucose oxidase (GOD) and horseradish peroxidase (HRP) were co-immobilized by microencapsulation in a sol-gel film derived from tetraethyl orthosilicate (TEOS The calibration plots for glucose were established by the optical fiber glucose sensor fabricated by attaching the bienzyme silica gel onto the glass window of the fiber bundle. The linear range was 0 2-2 mmol / L and the detection limit was approximately 0 12 mmol / L. The relative standard deviation was 5.3% (n = 6). The proposed biosensor was applied to glucose assay in of oxacin injection successfully.