论文部分内容阅读
Aflatoxins are the most potent spontaneous carcinogen known as secondary metabolite. Aflatoxins detection in peanut is challenging on account of inherent complexity of high fatty matrices. A modified Quick Easy Cheap Effective Rugged Safe(QuEChERS) method couple with liquid chromatograph-mass spectrometer(LC-MS/MS) has been developed to monitor aflatoxins in peanut. It?s a simple and rapid sample preparation for aflatoxins analysis compared with common method. Microporous metal-organic framework MIL-101(Cr) was used as absorbent in this research. Promising properties of high porosity and surface area endowed MIL-101(Cr) with efficiently aflatoxin enrichment with other matrix exclusion from high fatty sample. Negligible matrix effect was observed due to efficient extract clean up with MIL-101 sorbent. Recoveries ranged from 73.5% to 98.4% for aflatoxins, with relative standard deviations less than 7%. Limit of detection(LOD) and limit of quantitation(LOQ) for aflatoxins in peanut were 0.05-0.1 μg/kg and 0.2-0.6 μg/kg respectively. By using method described in this research, 15 real samples were successfully analyzed in which 6 samples were contaminated. MIL-101(Cr) was used to reduce matrix effect and develop a quick and effective method for aflatoxins detection in peanut via LC-MS/MS for the first time.