论文部分内容阅读
故障现象一在医院设备运行高峰期,正常扫描后的定位图像显示速度减慢且不完全。由此造成不能进行扫描前的准确定位。故障分析与检修初步分析可能为设备内网络连接系统故障。分别取下ICS侧交换机与IRS机柜中交换机、IRS机柜交换机与IRSTCR计算机连接网线,用网线测线器测试,显示
Fault phenomenon First, the peak of the hospital equipment operation, the normal scan of the positioning image display slow and not complete. As a result, accurate positioning before scanning can not be performed. Fault Analysis and Maintenance A preliminary analysis may be that the network connection system in the equipment is faulty. Remove the switches on the ICS side and the switches in the IRS cabinet respectively, and connect the IRS cabinet switch to the IRSTCR computer to test the network cable