论文部分内容阅读
目的观察全自动清洗消毒器对呼吸机外置回路的清洗消毒效果。方法采用目测法和细菌培养方法,对临床使用后的呼吸机外置回路清洗消毒效果进行了检测。结果经全自动清洗消毒器清洗消毒的316套呼吸机共计3792件外置回路,合格率达99.7%。结论使用全自动清洗消毒器清洗呼吸机外置回路不仅可保证清洗质量,亦可达到消毒效果要求,提高了工作效率。
Objective To observe the effect of automatic cleaning and disinfecting on the cleaning and disinfection of ventilator external circuit. Methods Visual method and bacterial culture method were used to detect the effect of external circuit cleaning and disinfection of ventilator after clinical use. The results of automatic cleaning and disinfection of the 316 sets of ventilators cleaning a total of 3792 external loop, with a pass rate of 99.7%. Conclusion The use of automatic cleaning and disinfecting external loop cleaning ventilator can not only ensure the quality of cleaning, disinfection can achieve the desired effect, improve work efficiency.