论文部分内容阅读
在使用岛津 ZS- 5 Y5 0 0 MA X光机当中遇到如下两例故障 :一是手动设置摄影条件时 ,k V及 S调节只增不减 ,其他正常。初次考虑为控制面板及相关电路问题 ,后查是由于连接机控制面板线太长 ,保养时将其拉断所致。另一故障为图像伪影 ,图像影一般为电视及摄像部分的问题 ,后查为
The use of Shimadzu ZS-5 Y5 0 0 MA X-ray machine which encountered the following two cases of failure: First, manually set the shooting conditions, k V and S adjustments increased only, the other normal. The first consideration for the control panel and related circuit problems, after check is due to the connector control panel line is too long, it will be pulled off when maintenance. Another problem is the image artifacts, the video image is generally part of the television and video problems, check after