论文部分内容阅读
随着乡镇卫生院经济不断发展,近年来不少乡镇卫生院已开始购置和使用高容量医用X线机。高容量新机型应用,其放射防护要求也相应提高。因此,为执行“三同时”制度,体现防护最优化原则,既避免重复性改建和经济浪费,又保证放射防护效果,是基层实际工作中需探讨和研究的课题。笔者在1998年审查监督过两个乡镇卫生院新购双床双球管医用X线机的安装,根据实际情况,提出两室安装的技术性监督意见,对此进行论证性监督监测,并与本市另两家基本条件相似的乡镇医院进行对比分析,认为此机型安装方式在基层卫生院切实可行。现将结
With the continuous development of township hospitals, many township hospitals have started to purchase and use high-capacity medical X-ray machines in recent years. High-capacity new models of applications, the radiation protection requirements also increased accordingly. Therefore, in order to implement the “three simultaneous” system and to embody the principle of optimal protection, we should avoid repetitive alterations and economic waste while ensuring the radiological protection effect. This is a subject to be explored and studied in the grassroots work. In 1998, the author reviewed and supervised the installation of newly purchased dual-bed double-tube medical X-ray machines in two township hospitals, and put forward the technical supervision opinions on the installation of two rooms according to the actual situation, The other two basic conditions similar to the township hospitals for comparative analysis that the installation of this model is feasible in primary health clinics. Now will end