Study on dosimetric and radiation safety performance parameters of helical tomotherapy unit

来源 :中华放射医学与防护杂志 | 被引量 : 0次 | 上传用户:g471151931
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
Objective

To understand the functional status and radiation protection requirements of helical tomotherapy (TOMO) unit when it is applied in clinic treatment so as to promote the construction of TOMO quality control system.

Methods

A helical tomotherapy unit with a type of tomotherapy Hi-Art was employed as the radiotherapy unit. A series of tests and analyses on dosimetric and radiation safety performance parameters of the TOMO unit were conducted with the use of a A1SL ionization chamber, a set of solid water equivalent phantoms and other test equipments. The performance parameters were compared with that of a conventional 6 MV accelerator with a type of ARTISTE.

Results

The deviations of static and dynamic output dose were -1.51% and 1.18%, respectively. The values of PDD10/PDD1.5 and PDD20/PDD1.5 were 0.608 and 0.318 within the range of X-ray quality parameters calibration. Under the different field conditions of 40 cm×5 cm, 40 cm×2.5 cm and 40 cm×1 cm, the stability of transverse/longitudinal dose distribution curve was no more than 1.20% and the offset distance of Jaw width was smaller than 0.5 mm. Radiation leakage ratio of the treatment rotating center from the multileaf collimator (MLC) was 0.20%. No matter the average or the maximum value of the leakage radiation from couch plane was less than 0.02%.

Conclusions

The performances and testing requirements of TOMO unit are different from a conventional electron linear accelerator, the quality control index of TOMO unit should be set seperately and tested strictly in terms of the specific requirements of the unit by complying with the QC programme. The reliable quality control system should be established and effectively implemented to serve the radiotherapy quality and radiation safety effectively in the tomotherapy.

其他文献
期刊
期刊
期刊
期刊
目的观察不同剂量X射线照射对小鼠辅助性T细胞相关细胞因子的影响。方法100只BALB /c小鼠用随机数字表法分为4组,包括健康对照组和X射线照射组(分别予2、4和6 Gy照射),每组25只。于照射前、照射后7、14、21和28 d,对小鼠一般情况进行观察,对外周血进行白细胞计数,用蛋白芯片检测血清中辅助性T细胞相关细胞因子含量,并用ELISA确证细胞因子的浓度变化。结果照射后,小鼠外周血白细胞14
目的研究自噬在照射致人食管鳞癌Eca-109细胞死亡过程中的作用。方法选用食管鳞癌Eca-109细胞,分为对照组、5 mmol/L给药组、10 mmol/L给药组、单纯照射组、照射+5 mmol/L组、照射+10 mmol/L组,共6组。采用Western blot检测不同处理组自噬标志物LC3B表达水平;GFP-LC3转染食管鳞癌Eca-109细胞后监测自噬体数量变化;MTT法检测不同处理组细胞
目的 探讨胸主动脉CT血管造影(CTA)成像时的适应性统计迭代重建(ASIR)的最佳权重和优化低剂量扫描参数.方法 将体质量指数为19~24 kg/m2的120例患者随机数字表法分成6组,用宝石能谱CT进行胸主动脉CTA扫描,扫描范围为290 ~ 330 mm,第1组扫描条件为机器默认的100 kV、240 mAs,使用10%~100%权重的适应性统计迭代重建,对图像进行重建并计算信噪比(SNR)
2014年9月26日,中国科技论文统计结果发布,《中华放射医学与防护杂志》的影响因子继续保持增长的势头,综合评价总分在军事医学与特种医学类排名第一,在1989种中国科技核心期刊中名列第128位。为此,入选“第3届中国精品科技期刊”(Outstanding S&T Journals of China),“中国精品科技期刊顶尖学术论文(F5000)”项目来源期刊。
期刊
期刊
目的探讨12C重离子束对人淋巴细胞增殖以及周期、凋亡的影响。方法12C重离子束照射人淋巴细胞Peng-EBV,吸收剂量分别为0(对照组)、0.5、2.0 Gy。照射后用MTS法检测细胞增殖活力,流式细胞仪检测细胞周期和细胞凋亡。结果与对照组相比,0.5 Gy照射可以增加细胞增殖活力(t=2.66~14.45, P<0.05),而2.0 Gy照射降低了细胞活力(t=7.65~64.45, P<0.0