论文部分内容阅读
动式吸入染毒装置,是研究分布于空气中的化学物质以吸入方式进行染毒所必需的实验设备。我室为开展工业毒物中毒防治的实验研究,自1972年起,参考了国内外有关资料,从我国的实际出发,自行设计和研制了一套简易的动式吸入染毒装置。该染毒装置主要由两个有机玻璃动式染毒柜、空气调节净化系统和毒物浓度控制系统三部分组成。 设计与构造 (一)染毒室与染毒装置布置 染毒室分为空调室和控制室两部分。空调室内装设窗式空调器、轴流式事故排风扇,染毒柜、通风管道和净化系统。配电
Dynamic inhalation and exposure device, is to study the chemical substances distributed in the air by inhalation of the necessary experimental equipment exposure. Our laboratory has conducted experimental research on the prevention and control of industrial poisoning poisoning. Since 1972, with reference to the relevant information at home and abroad, a set of simple dynamic inhalation and poisoning devices has been designed and developed by our country based on the actual conditions in our country. The device is mainly composed of two plexiglass dynamic poisoning cabinets, air conditioning purification system and poison concentration control system composed of three parts. Design and construction (A) of the poisoning room and the device is equipped with poisoning room into the air-conditioned room and control room in two parts. Air conditioners are equipped with window air conditioners, axial fans, poison cabinets, ventilation ducts and decontamination systems. power distribution