伪故障三例

来源 :医疗装备 | 被引量 : 0次 | 上传用户:zbgqx123456
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
例一:用户反应一台开放式新生儿抢救台超温报警。分析检修:(1)该机工作温度设定在26℃,而显示屏上为283℃。此机在偏离设定温度2℃就会自动报警,根据现象可判断报警部分无故障。(2)关机后,取下传感器使它接近冷热源,测量其直流电阻在90Ω左右变化,用同型传感器比较俩者无差别,? Example 1: The user responds to an open neonatal rescue station over-temperature alarm. Analysis and maintenance: (1) The working temperature of the machine is set at 26°C, and the display is 283°C. This machine will automatically alarm when it deviates from the set temperature by 2°C. According to the phenomenon, it can be judged that there is no fault in the alarm part. (2) After shutting down, take off the sensor to make it close to the cold and heat source, measure the DC resistance changes around 90Ω, and compare them with the same sensor without any difference.
其他文献
ZFS302-Ⅰ型中频X线机是北京万东医疗装备公司于九五年研制生产的用8098单片机芯控制的数控X线机.由于电源采用了三项十二波整流,kV、mA采用闭环采样控制等一系列较先进的技
我院28年间共收治急性心肌梗塞76例,并发心律失常53例,总检出率为69.7%。其中室性心律失常39例,房性心律失常24例,传导障碍17例.后、下壁梗塞者缓慢性心律失常、房室传导阴滞
近年,国内外大量的动物和临床研究表明,非创伤性经食道心房起搏术方法可靠,安全简便,可重复应用,也适用于基层医疗单位。现将经食道心房调搏的基本方法和临床应用作一初步介
故障现象:用校验带测试工作正常,但用尿样带测试出现TROUBLE-4错误代码。分析检修:根据故障现象判别,用校验带测试正常,说明仪器控制部分、光学系统、扫描机构、V/F转换器、CPU,等工作正
由于电子管型交流稳压电源具有功率大,电气性能稳定,调压能力强,稳压精度高、故障率低等优点,故在我国医院医疗装备中,占有比例较大,成为医院基础装备的设备之一。为此,我们
故障现象一:该机只能做头部和腰椎扫描,而体部病人不能扫描,显示屏提示:1,Mcuerror#2002Scannernotincworccwposition2,Mcuerror#4769MetatimeoutonendofBeammotion故障分析及排除:从以上故障现象和错 Symptom 1: The machine can only do head and l
故障现象:透视正常,点片时ready灯亮,且传片曝光程序动作正常,所点片子是白板,只在右下角病人卡处有黑化度记录。故障分析与检修:1排除洗片机故障在其它机器上所点片子冲洗后正常,
F - 8 2 0血球计数仪常规故障为小孔堵塞 ,及时快速地处理好此类故障不仅节约了时间 ,节约了资金 ,而且保障了仪器的正常使用并提高了测量的准确度。本文详尽地介绍了小孔堵
F108—V型遥控透视X线机是由北京X线机厂生产,为组合式全波整流,全封闭式X射线管头,配有日本东芝高分辨影像增强器和航天工业部研制的电视系统。现将在使用过程中出现的典型故障介
BJ—6B直线加速器具有自动楔形系统,它的60度楔形板在所给信号的控制下,自动进入或退出照射野。根据WEDGE野和OPEN野照射剂量的不同比例分配,可以合成60—0度之间的任何楔形板。OPEN野时60度