本文将意大利IOTA型三坐标测量机两个典型故障的排除方法作一简单介绍。一、整机不能启动 W_A交流电源单元出了故障。检查故障时,先卸掉W_A单元后面板上的大电缆,从电气柜中将W_A整个箱体拉出,打开屏蔽盒,在检查主要元件的基础上,用外接220和24电源的方法,进行通电检查。若继电器TL_1有毛病,可先用细砂纸和酒精擦洗继电器触点。如不行,需卸下重新调节弹力,反复试验调整间隙后便可恢复工作。二、Z轴数显表1至5位计数模糊不清故障出在某个MC计数单元上。该仪器
This article will Italy IOTA CMM two typical troubleshooting methods for a brief introduction. First, the whole machine can not start W_A AC power unit out of trouble. To check the fault, first remove the large cable on the rear panel of the W_A unit, pull the entire W_A box out of the electrical cabinet, and open the shield box. On the basis of checking the main components, use external 220 and 24 power supply Power check. If relay TL_1 is faulty, scrub the relay contacts with fine sandpaper and alcohol. If not, remove the need to re-adjust the elastic force, after repeated trials to adjust the gap can resume work. Second, the Z-axis digital display 1 to 5-bit count fuzzy fault out of a MC count unit. The instrument