论文部分内容阅读
在氨性介质中,邻菲啰啉与Cr(Ⅵ)有一峰电流较大的催化波,Cr(Ⅵ)浓度在0.006~0.16μg/ml范围内与峰电流呈线性关系,检测限为7.7×10~(-8)mol/L,该法可用于矿石和水样中Cr的测定。