例1 故障现象:漏气分析与检修:当向气袋内加压以后,发现汞柱自然回落,则可确定为漏气,然后用手捏紧皮球与气袋的连接管.如果汞柱仍在下降,则是气袋破孔漏气,反之是皮球漏气。对于前者,应换新品,或可以进行修补。皮球漏气,一般是皮球破损老化引起,有时又跟前气阀内小橡皮管两切口密合处有絮状物存在有关,可分别对症处理。例2 故障现象:皮球后气阀跑气
Example 1 Symptom: Leakage Analysis and Maintenance: When the pressure inside the airbag and found that the natural fall of mercury, you can determine the leakage, and then hand pinch the ball and the air bag connection tube. If the mercury is still In descent, it is the airbag break hole leakage, on the contrary is the ball leak. For the former, should be new, or can be repaired. The ball leaks, the ball is generally caused by damage to the aging of the ball, sometimes with the small rubber tube in front of the valve two incision at the junction of the existence of flocs can be symptomatic treatment. Example 2 Symptoms: the ball after the gas valve run gas