本文以用于红外探测器的基本特性探测率D参数测试的辐射正弦调制器的设计式作推导,给出五个正弦调制器设计数据表。一、物理概念红外探测器D参数测试系统均设选频放大器,选频放大器要求系统中的黑体辐射应经正弦调制,以让探测器的光电转换信号获得最大通过,这样的测量才精确。常用的圆光阑调制器不能满足辐射正弦调制的要求,因该调制器为非正弦调制器,其调制特性无法改变。文献在对圆光阑调制器的非正弦调制特性作分析时,作者均提到“理想的”正弦调
In this paper, we use the design of radiation sine modulator for testing the basic characteristic detection rate D of infrared detectors, and give five sine modulator design data tables. First, the physical concept Infrared detector D parameter test system is set to select the frequency amplifier, the frequency selective amplifier requires blackbody radiation in the system should be sinusoidal modulation, so that the photoelectric conversion of the detector to maximize the adoption of such measurements are accurate. Commonly used circular aperture modulator can not meet the requirements of radiation sine modulation, because the modulator is a non-sinusoidal modulator, the modulation characteristics can not be changed. In the analysis of the non-sinusoidal modulation characteristics of a circular aperture modulator, the author mentions “ideal” sinusoidal modulation