正电子发射断层成像(Positron Emission Tomography,PET)作为最灵敏和具有定量测量能力的功能分子影像技术,越来越广泛地应用于生物医学研究,如疾病的动物模型、新药物的研发和新治疗方法的评估等。提高探测器的性能是改进PET仪器性能的关键,PET探测器通常由分割的闪烁晶体阵列和光探测器组成,文中使用位置灵敏光电倍增管和不同晶体表面特性的硅酸钇镥((Lu,Y)2Si O5,LYSO)晶体阵列,对新型的双端读出三维PET探测器和传统的单端读出二维PET探测器的性能进行了测量。实验结果表明,对于双端读出PET探测器,两种晶体阵列提供相近的晶体分辨图和能量分辨率,但非抛光晶体阵列提供好的深度分辨率,双端读出PET探测器需要使用表面不抛光的晶体阵列;对单端读出PET探测器,抛光晶体阵列提供好的晶体分布图和能量分辨率,单端读出PET探测器需要使用表面抛光的晶体阵列。
Positron Emission Tomography (PET), the most sensitive and quantitative molecular imaging technique, is increasingly used in biomedical research such as animal models of disease, development and treatment of new drugs Method of assessment and so on. To improve the performance of the detector is the key to improve the performance of the PET apparatus. The PET detector usually consists of a split scintillation crystal array and a photodetector. In this paper, position-sensitive photomultiplier tube and yttrium silicate with different crystal surface properties ((Lu, Y ) 2Si O5, LYSO) crystal arrays were used to measure the performance of the new double-ended readout 3D PET detector and the conventional single-ended readout 2D PET detector. Experimental results show that the two crystal arrays provide similar crystal resolution and energy resolution for a two-terminal readout PET detector, but a non-polished crystal array provides good depth resolution and a two-terminal readout PET detector requires the use of a surface Non-polished crystal arrays; good single-end read PET detectors, polished crystal arrays provide good crystal distribution and energy resolution, single-ended read PET detectors require the use of surface-polished crystal arrays.