本文介绍用2,3—二氨基萘于血、发、尿及动物组织中硒含量的萤光测定法,此法最低检出浓度为0.001μg/ml,最低检出量为3ng。用10—500ng硒的标准溶液作重复测定,其检出值的变异系数为7.3—2.4%;血、发、尿及动物组织中硒检出值的变异系数分别为3.9、5.5、3.3及5.6%;回收率依次分别为97.0±2.0,95.0±1.9,97.8±1.1及99.8±1.8%。用不等量以上样品检出的硒含量近似,这说明萃取物中的硒未受到外来物质的干扰。本法的精密度及准确度均高。
This article describes the use of 2,3 - diaminonaphthalene in blood, hair, urine and animal tissue fluorescence of selenium content, the minimum detectable concentration of 0.001μg / ml, the minimum detectable amount of 3ng. With 10-500ng selenium standard solution for repeated determination, the detection coefficient of variation coefficient of 7.3-2.4%; blood, hair, urine and animal tissue selenium detection value of the coefficient of variation were 3.9,5.5,3.3 and 5.6 %; The recoveries were 97.0 ± 2.0, 95.0 ± 1.9, 97.8 ± 1.1 and 99.8 ± 1.8% respectively. The selenium content detected by the unequal amount of samples above shows that the selenium in the extract is not disturbed by foreign substances. The precision and accuracy of this Law are high.