高纯金属铈的ICP-AES分析和稀土元素间的光谱干扰及其校正

来源 :光谱学与光谱分析 | 被引量 : 0次 | 上传用户:huapdf
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
本文采用乙醇预去溶方式进样,ICP-AES直接同时测定了>99.99%的高纯金属铈中5个痕量稀土杂质元素,并讨论了基体量对检测限的影响,稀土元素间的光谱干扰及其校正方法等。当样品溶液中铈的浓度为5mg/ml时,铈中稀土杂质元素的测定下限为:镧、钕和钇0.001%,镨和钐0.003%。其相对标准偏差为1.9~6.4%。 In this paper, ethanol was used to pre-dissolve the sample and ICP-AES was used to simultaneously determine five trace rare earth elements in> 99.99% high purity cerium. The influence of substrate amount on the detection limit, Interference and its correction methods. When the concentration of cerium in the sample solution is 5 mg / ml, the lower limit of determination of the rare earth impurity element in cerium is 0.001% of lanthanum, neodymium and yttrium, 0.003% of praseodymium and samarium. The relative standard deviation of 1.9 ~ 6.4%.