X380—D型洗片机故障检修1例

来源 :实用医技杂志 | 被引量 : 0次 | 上传用户:QCLHQCLH
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
1.故障现象 开机后定影槽内液体不循环。定影液不循环,可能有两方面原因所致。一方面是因循环管堵塞或充满气体;另一方面则为循环泵所致。首先由易向难检查循环管,发现正常,则故障为循环泵及其电路所致。查循环泵输入电压正常。将循环泵拆下,在静态下查其绕组及 1. Symptom The fluid in the fixing tank does not circulate after the power is turned on. The fixing fluid does not circulate. There may be two reasons. On the one hand, it is due to blockage of the circulation pipe or full of gas; on the other hand, it is due to the circulation pump. First of all, it is difficult to check the circulation pipe from the easy to find that it is normal, then the fault is caused by the circulation pump and its circuit. Check the circulating pump input voltage is normal. Remove the circulation pump, statically check its windings and